Biomechanics of Bone / Bone Biomechanics
Mohammad Nikkhoo; Ali Tahassori; Mohammad Haghpanahi
Volume 8, Issue 3 , September 2014, , Pages 203-212
Abstract
To develop the advanced technologies in medical device industry, design and manufacturing of cervical cage was performed in Iran for the first time. This research-based industrial project should be accomplished based on precise biomechanical studies and mechanical tests. Hence, this study presents the ...
Read More
To develop the advanced technologies in medical device industry, design and manufacturing of cervical cage was performed in Iran for the first time. This research-based industrial project should be accomplished based on precise biomechanical studies and mechanical tests. Hence, this study presents the optimization and biomechanical functional investigations of the first Iranian cervical cage (Manufactured by Attila Ortopaed Co.). For this purpose the intact cervical spine (C2-C7) was developed and was validated with in-vitro experiments. Three inputs (i.e. geometrical parameters of the cage) and two outputs (i.e. deformation of the teeth in static and dynamic tests) parameters were selected for optimization procedure. Furthermore, the surgery in C5-C6 level was simulated by implanting the cervical cage. Finally, the biomechanical responses were investigated. The result confirmed that the biomechanical response of cervical cage is within the standard range and can be used well in clinics for surgical procedures.